Specification:
Fluke 718 Series | |
Measure pressure | (Internal sensor) |
Accuracy | 0.025% full scale |
Functions | Zero, min/max, hold, damp |
Dry air, non-corrosives only | |
Fluke 718 1G | |
Range | -1 psi to +1 psi, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa) |
Resolution | 0.0001 psi, 0.001 mbar, 0.0001 kPa |
Over pressure | 5x Full Scale |
Functions | Zero, Min, Max, Hold, Damp |
Fluke 718 30G | |
Range | -12 psi to 30 psi, (-850 mbar to 2 bar, -85 to 206.84 kPa) |
Resolution | 0.001 psi, 0.1 mbar, 0.01 kPa |
Over pressure | 5x Full Scale |
Functions | Zero, Min, Max, Hold, Damp |
Fluke 718 100G | |
Range | -12 psi to 100 psi, (-850 mbar to 7 bar, -85 to 689.48 kPa) |
Resolution | 0.01 psi, 0.1 mbar, 0.01 kPa |
Over pressure | 2x Full Scale |
Functions | Zero, Min, Max, Hold, Damp |
Fluke 718 300G | |
Range | -12 psi to 300 psi, (-850 mbar to 20 bar, -85 to 2068.42 kPa) |
Resolution | 0.01 psi, 1 mbar, 0.1 kPa |
Over pressure | 375 psi 25 bar |
Functions | Zero, Min, Max, Hold, Damp |
Fluke 718 series via 700 Series Pressure Modules | |
Range | 29 pressure modules, 0 – 1 inH20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs |
Resolution | Per pressure module specs |
Accuracy | To 0.025% of full span, per pressure module specs |
Functions | Zero, Min, Max, Hold, Damp |
Media compatibility per pressure module specs | |
Fluke 718 Series Via Built in Pump | |
Range | -12 psi or -850 mbar to full scale Supported Pressure Units; psi, inH20 (4°C), inH20 (20°C), cmH20 (4°C), cmH20 (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm |
Loop power range | 24 V DC |
Accuracy | ±10% |
Drive; 20 mA into 1000 Ω for battery > 6.8 V; 700 Ω for battery 5.8 to 6.8 V | |
mA Measure | |
Range | 0 mA to 24 mA |
Resolution | 0.001 mA |
Accuracy | 0.015% + 1 count |
General Specifications | |
Maximum voltage | 30 volts (non-operating) |
Temperature | -40°C to 60°C |
Operating temperature | -10°C to 55°C |
Relative humidity | 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C) |
Operating altitude | 3,000 M max |
Shock | 1 M drop test |
Vibration | Random, 2 g, 5 Hz – 500 Hz |
Safety | CSA C22.2 No. 1010.1: 1992 |
EMC | EN50082-1: 1992 and EN55022: 1994 Class B |
Size | 216 x 94 x 66 mm (8.50 x 3.72 x 2.60 in) |
Weight | 992 g (35 oz) |
Power, battery life | 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718 |
Battery life | 4 to 20 hours typical, depending on functions used |
Warranty | Three years (one year on pressure pump in Fluke 718) |
Display | LCD, 5 digit pressure and current simultaneous |