Website

www.canelectronic.com

Rabu, 14 September 2016

Pressure Calibrator Fluke 718-30G

Fluke

Pressure Calibrator 718-30G



CAN Electronic Indonesia
www.canelectronic.com
canelectronic.ind@gmail.com
021 6232 0266, 2607 1120 / 26





Specification:


Fluke 718 Series
Measure pressure(Internal sensor)
Accuracy0.025% full scale
FunctionsZero, min/max, hold, damp
Dry air, non-corrosives only
Fluke 718 1G
Range-1 psi to +1 psi, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa)
Resolution0.0001 psi, 0.001 mbar, 0.0001 kPa
Over pressure5x Full Scale
FunctionsZero, Min, Max, Hold, Damp
Fluke 718 30G
Range-12 psi to 30 psi, (-850 mbar to 2 bar, -85 to 206.84 kPa)
Resolution0.001 psi, 0.1 mbar, 0.01 kPa
Over pressure5x Full Scale
FunctionsZero, Min, Max, Hold, Damp
Fluke 718 100G
Range-12 psi to 100 psi, (-850 mbar to 7 bar, -85 to 689.48 kPa)
Resolution0.01 psi, 0.1 mbar, 0.01 kPa
Over pressure2x Full Scale
FunctionsZero, Min, Max, Hold, Damp
Fluke 718 300G
Range-12 psi to 300 psi, (-850 mbar to 20 bar, -85 to 2068.42 kPa)
Resolution0.01 psi, 1 mbar, 0.1 kPa
Over pressure375 psi 25 bar
FunctionsZero, Min, Max, Hold, Damp
Fluke 718 series via 700 Series Pressure Modules
Range29 pressure modules, 0 – 1 inH20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs
ResolutionPer pressure module specs
AccuracyTo 0.025% of full span, per pressure module specs
FunctionsZero, Min, Max, Hold, Damp
Media compatibility per pressure module specs
Fluke 718 Series Via Built in Pump
Range-12 psi or -850 mbar to full scale Supported Pressure Units; psi, inH20 (4°C), inH20 (20°C), cmH20 (4°C), cmH20 (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop power range24 V DC
Accuracy±10%
Drive; 20 mA into 1000 Ω for battery > 6.8 V; 700 Ω for battery 5.8 to 6.8 V
mA Measure
Range0 mA to 24 mA
Resolution0.001 mA
Accuracy0.015% + 1 count
General Specifications
Maximum voltage30 volts (non-operating)
Temperature-40°C to 60°C
Operating temperature-10°C to 55°C
Relative humidity95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude3,000 M max
Shock1 M drop test
VibrationRandom, 2 g, 5 Hz – 500 Hz
SafetyCSA C22.2 No. 1010.1: 1992
EMCEN50082-1: 1992 and EN55022: 1994 Class B
Size216 x 94 x 66 mm (8.50 x 3.72 x 2.60 in)
Weight992 g (35 oz)
Power, battery life9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Battery life4 to 20 hours typical, depending on functions used
WarrantyThree years (one year on pressure pump in Fluke 718)
DisplayLCD, 5 digit pressure and current simultaneous


Tidak ada komentar: