Website

www.canelectronic.com

Rabu, 14 September 2016

Documenting Process Calibrator Fluke 754

Fluke

Documenting Process Calibrator-HART 754



CAN Electronic Indonesia
www.canelectronic.com
canelectronic.ind@gmail.com
021 6232 0266, 2607 1120 / 26




Specification:


Measurement Accuracy
Voltage DCRange/Resolution1 Year2 Years
100.000 mV0.02% + 0.005 mV0.03% + 0.005 mV
3.00000 V0.02% + 0.00005 V0.03% + 0.00005 V
30.0000 V0.02% + 0.0005 V0.03% + 0.0005 V
300.00 V0.05% + 0.05 V0.07% + 0.05 V
Voltage AC3.000 V (40 Hz to 500 Hz) / 0.001 V0.5% + 0.002 V1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V0.5% + 0.02 V1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V0.5% + 0.2 V1.0% + 0.2 V
Current DC30.000 mA0.01% + 5 uA0.015% + 7 uA
110.00 mA0.01% + 20 uA0.015% + 30 uA
Resistance10.000 Ω0.05% + 50 mΩ0.07% + 70 mΩ
100.00 Ω0.05% + 50 mΩ0.07% + 70 mΩ
1.0000 kΩ0.05% + 500 mΩ0.07% + 0.5 Ω
10.000 kΩ0.1% + 10 Ω0.15% + 15 Ω
Frequency1.00 to 110.00 Hz / 0.01 Hz0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz0.05 kHz
Source Accuracy
1 Year2 Years
Voltage DC100.000 mV0.01% + 0.005 mV0.015% + 0.005 mV
1.00000 V0.01% + 0.00005 V0.015% + 0.0005 V
15.0000 V0.01% + 0.0005 V0.015% + 0.0005 V
Current DC22.000 mA (source)0.01% + 0.003 mA0.02% + 0.003 mA
Current sink (simulate)0.02% + 0.007 mA0.04% + 0.007 mA
Resistance10.000 Ω0.01% + 10 mΩ0.015% + 15 mΩ
100.00 Ω0.01% + 20 mΩ0.015% + 30 mΩ
1.0000 kΩ0.02% + 0.2 Ω0.03% + 0.3 Ω
10.000 kΩ0.02% + 3 Ω0.03% + 5 Ω
Frequency0.1 to 10.99 Hz0.01 Hz
0.01 to 10.99 Hz0.01 Hz
11.00 to 109.99 Hz0.1 Hz
110.0 to 1099.9 Hz0.1 Hz
1.100 to 21.999 kHz0.002 kHz
22.000 to 50.000 kHz0.005 kHz
Technical Data
Data log functionsMeasure functionsVoltage, current, resistance, frequency, temperature, pressure
Reading rate1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length8000 readings (7980 for 30 or 60 readings/minute)
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
Rate4 steps/second
Trip detectContinuity or voltage (continuity detection not available when sourcing current)
Loop power functionVoltageSelectable, 26 V
Accuracy10%, 18 V minimum at 22 mA
Maximum current25 mA, short circuit protected
Maximum input voltage50 V DC
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Manual stepSelectable step, change with arrow buttons
AutostepFully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature-10°C to +50°C
Storage temperature-20°C to +60°C
Dust/water resistanceMeets IP52, IEC 529
Operating altitude3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvalsCAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight1.2 kg (2.7 lb)
BatteriesInternal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Battery life>8 hours typical
Battery replacementReplace without opening calibrator; no tools required
Side port connectionsPressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity1 week of calibration procedures results
90 day specificationsThe standard specification interval for the 750 Seriesare 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Temperature, Resistance Temperature Detectors
Degrees or % of reading - Type (α)Range °CMeasure °C1
1 year2 year
100 Ω Pt (385)-200 to 100
100 to 800
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
200 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
500 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
1000 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3916)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3926)-200 to 100
100 to 630
0.08°C
0.02% + 0.06°C
0.16°C
0.04% + 0.12°C
10 Ω Cu (427)-100 to 2600.2°C0.4°C
120 Ω Ni (672)-80 to 2600.1°C0.2°C
Source currentSource °CAllowable current2
1 year2 year
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
500 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
250 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
150 μA0.06 C
0.017% + 0.05°C
0.12 C
0.034% + 0.10°C
0.1 mA to 1 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
3 mA0.2°C0.4°C0.1 mA to 10 mA
1 mA0.04°C0.08°C0.1 mA to 10 mA
1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples
TypeSource °CMeasure °CSource °C
1 year2 years1 year2 years
E-250 to -2001.32.00.60.9
-200 to -1000.50.80.30.4
-100 to 6000.30.40.30.4
600 to 10000.40.60.20.3
N-200 to -1001.01.50.60.9
-100 to 9000.50.80.50.8
900 to 13000.60.90.30.4
J-210 to -1000.60.90.30.4
-100 to 8000.30.40.20.3
800 to 12000.50.80.30.3
K-200 to -1000.71.00.40.6
-100 to 4000.30.40.30.4
400 to 12000.50.80.30.4
1200 to 13720.71.00.30.4
T-250 to -2001.72.50.91.4
-200 to 00.60.90.40.6
0 to 4000.30.40.30.4
B600 to 8001.32.01.01.5
800 to 10001.01.50.81.2
1000 to 18200.91.30.81.2
R-20 to 02.32.81.21.8
0 to 1001.52.21.11.7
100 to 17671.01.50.91.4
S-20 to 02.32.81.21.8
0 to 2001.52.11.11.7
200 to 14000.91.40.91.4
1400 to 17671.11.71.01.5
C0 to 8000.60.90.60.9
800 to 12000.81.20.71.0
1200 to 18001.11.60.91.4
1800 to 23162.03.01.32.0
L-200 to -1000.60.90.30.4
-100 to 8000.30.40.20.3
800 to 9000.50.80.20.3
U-200 to 00.60.90.40.6
0 to 6000.30.40.30.4
BP0 to 10001.01.50.40.6
1000 to 20001.62.40.60.9
2000 to 25002.03.00.81.2
XK-200 to 3000.20.30.20.5
300 to 8000.40.60.30.6

Tidak ada komentar: